Data Sheet
| Temperature
Range |
-73°C
to +175°C |
| Control
Tolerance |
±0.5°C
(±0.2°C Typical) (Measured at the control sensor
after stabilization) |
| Uniformity |
±1.0°C
(±0.5°C Typical) (Variations throughout the chamber
after stabilization) |
| Heat
Up Transition Time* |
5°C/minute
typical |
| Cool
Down Transition Time* |
| Start
Temp |
End
Temp |
| +23°C |
0°C |
-40°C |
-55°C |
-65°C |
-73°C |
| +23°C |
----- |
5
min |
20
min |
30
min |
40
min |
Ultimate |
| +85°C |
16
min |
25
min |
45
min |
57
min |
69
min |
Ultimate |
Rate
Of Change
To calculate rate of change for a particular condition, take the
difference between the Start Temp and End Temp and divide by the
Transition Time.
Cool Down Example: From +85°C to -40°C = 125°C
/ 45 min = 2.78°C/min. |
| *Note: Transition
times are measured after a 1 hour soak at the respective start
temperature with an empty chamber, as indicated on the temperature
controller, 23°C ambient. |
| Live
Load Capacity |
| +23°C |
0°C |
-40°C |
-55°C |
-65°C |
| 300 Watts |
255 Watts |
160 Watts |
110 Watts |
80 Watts |
| Refrigeration
and Heating System |
| High
Stage Refrigerant |
R-404A
(Dupont HP-62) |
| Low
Stage Refrigerant |
R-508B
(Dupont SUVA-95) |
| Compressors |
1/2
HP x 1/3 HP Copeland hermetic compressors in a cascade configuration. |
| Condenser |
Air
Cooled |
| Heater
Power |
600
Watts |
| Instrumentation |
| Temperature
Controller |
256
step, 40 profile, ramp and soak programmable memory. RS-232/485
interface.
More details >> |
|
| Limit
Controller |
Independent
high and low temperature limits. Triggers an audible alarm and
shuts down the chamber. |
| Power
Requirements |
| Input Voltage |
120 V nominal
(110 to 126 VAC), 60 Hz, 1 PH
Max Current Draw 15 A, Recommended Minimum Service 20 A
Model 115-EX
export version available with 230V, 50Hz, 8A input |
Power Cord
and Plug |
| 6' Power
cord supplied with a molded NEMA 5-20P plug. |
 |
Plugs
into a standard NEMA 5-20R receptacle.
Use of an extension cord is not recommended. |
 |
Model 115-EX
export version is supplied with power cord for destination country. |
| Physical
Characteristics |
| Inside
Dimensions |
16" W
x 12" H x 14" D (1.55 Cubic feet)
406 mm W x 305 mm H x 353 mm D (44 liters) |
| Outside
Dimensions |
Benchtop
Model 115B
24" W x 44" H x 26" D (nominal)
610 mm W x 1118 mm H x 660 mm D
Floor Model
115F
24" W x 61" H x 26" D (nominal)
610 mm W x 1549 mm H x 660 mm D |
| Minimum
Installed Clearance |
6" from
the left and right side
12" from the rear |
| Window
Viewing Area |
8"
W x 6" H |
| Access
Ports |
All
new 115 chambers now include 3" ports (2.83" inside diameter)
on left and right side (two total). Supplied with silicone foam
port plugs |
| Weight |
Benchtop
Model 115B
Chamber Weight: 275 pounds
Shipping Weight: 320 pounds
Floor Model
115F
Chamber Weight: 300 pounds
Shipping Weight: 345 pounds |
| Sound
Level |
52
dBA in cooling mode (A-weighted, measured 36" from the front
surface, 63" from the floor, in a free-standing environment) |
| CE
Mark |
Export
version, Model 115-EX is CE Marked |
| NOTE: Performance
is typical and based on operation at 23°C (73°F) ambient
and nominal input voltage. Designed for use in a normal conditioned
laboratory. Operation at higher ambient temperatures will result
in decreased cooling performance. Low end limit derates to -38°C
when operating above 27°C (80°F) ambient. Operation above
30°C (85°F) or below 16°C (60°F) ambient is not
recommended.
|
|