   TestEquity Model 1007C Temperature Chamber - Detailed Specifications


Products

Why Buy a
TestEquity
Chamber

About
TestEquity

Contact Us

Home

<< Back to Product Page  PDF Data Sheet

Temperature Range -73°C to +175°C
Control Tolerance ±0.2°C (Measured at the control sensor after stabilization)
Uniformity ±0.5°C (Variations throughout the chamber after stabilization)

Cool Down Transition Time*

Start Temp

End Temp

+23°C 0°C -40°C -55°C -65°C -73°C
+23°C ----- 4 min 18 min 25 min 33 min Ultimate
+50°C 5 min 11 min 27 min 36 min 45 min Ultimate
+85°C 13 min 20 min 37 min 47 min 55 min Ultimate
+150°C 32 min 39 min 58 min 65 min 76 min Ultimate
 

Heat Up Transition Time*

Start Temp

End Temp

+23°C +50°C +85°C +125°C +150°C +175°C
+23°C ----- 1.5 min 7 min 14 min 20 min Ultimate
0°C 1.5 min 3.5 min 13 min 20 min 23 min Ultimate
-40°C 6 min 11 min 17 min 24 min 30 min Ultimate
-55°C 8 min 13 min 19 min 26 min 32 min Ultimate
-65°C 10 min 14 min 21 min 28 min 34 min Ultimate


Rate Of Change

To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example: From +50°C to -40°C = 90°C / 27 min = 3.33°C/min.
Heat Up Example: From -40°C to +50°C = 90°C / 11 min = 8.18°C/min.

*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber.


Live Load Capacity
+23°C 0°C -40°C -55°C -65°C
1,000 Watts 800 Watts 500 Watts 400 Watts 300 Watts

Refrigeration and Heating System

High Stage Refrigerant R-404A (Dupont HP-62)
Low Stage Refrigerant R-508B (Dupont SUVA-95)
Compressors 1 HP x 1 HP Copeland hermetic compressors in a cascade configuration.
Condenser Air Cooled
Heater Power 2,000 Watts
 

Instrumentation

Temperature Controller

256 step, 40 profile, ramp and soak programmable memory. RS-232/485 interface.
More details >>

Limit Controller Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample.
Chart Recorder   (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door.

Input Power Requirements

  Phase Current Draw Minimum Service
230 V ±10%, 60 Hz 1 PH 25 A 30 A
3 PH 19 A 25 A
 
208 V -5/+10%, 60 Hz 1 PH 28 A 35 A
3 PH 21 A 30 A

Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semi-balanced. Call for 50 Hz operation. Cooling performance is reduced 17% at 50 Hz.

Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.


Physical Characteristics

Inside Dimensions

24" W x 21" H x 24" D (7 cubic feet)

Outside Dimensions* 33" W x 67.75" H x 54.25" D (nominal)
* Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator motor and housing adds 6" to height.
Minimum Installed Clearance 18" from the left and right side
24" from the rear
Window Viewing Area 13.5" W x 11.5" H
Access Ports 4" Port on left and right side (two total)
Supplied with foam plugs
Weight Chamber Weight: 850 pounds
Shipping Weight: 1,000 pounds
Sound Level 62 dBA in cooling mode (A-weighted, measured 36" from the front or side surface, 63" from the floor, in a free-standing environment)

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.