| TECHNICAL
FEATURES
No-Sweat Cabinet
We use a combination of structural urethane foam and fiberglass
blanket insulation for a higher combined R-factor than just
fiberglass alone. The door sealing surface is kept warm by
circulating excess heat from the refrigeration system. The
result is a cabinet that does not sweat due to condensation
during low temperature operation.
The window
is constructed of 6 panes of tempered glass separated by low-conductivity
spacers to keep thermal losses to a minimum. Our unique design
virtually eliminates exterior window sweating during normal temperature
cycling conditions.
The stainless-steel
interior liner is heliarc welded, while the refrigeration compartment
behind the liner is double-insulated and carefully sealed. These
design features reduce thermal losses and prevent moisture migration,
ensuring years of optimum performance while keeping the cabinet
rust-free.
Reliability & Safety
Thats Designed In
Non-CFC refrigerants
are used in a cascade (two compressors) configuration. Expansion
valves provide optimum performance throughout the operating range.
Pressure regulators keep the operating pressures within safe limits
regardless of load conditions. Liquid injection ensures cool compressor
operation during high-temperature pull down for long life. The
low-stage desuperheater has its own fan, reducing load on the cascade
condenser without robbing capacity from the high-stage condenser
fan. Sequential starting of each compressor reduces the current
demand on start-up. Each compressor is protected by high/low pressure
switches. Care has been taken to use a minimum number of fittings,
joints and welds to reduce the possibility of refrigerant leaks.
All electrical
control components are UL approved. Each branch circuit is individually
fused. A main power disconnect switch is provided on the front
panel. The nichrome air-heater has a low watt-density and even
heat distribution for reliable performance and fast response.
A fusible link provides fail-safe protection against thermal
runaways, in addition to the microprocessor-based high/low limit
controller.
The side and
rear panels are removable to permit complete access to the refrigeration
system for ease of service and preventive maintenance. Shown
above with the optional circular chart recorder, which mounts
conveniently in the front. |